Blender
Air-Oxygen
COMPONENT DESCRIPTION
ITEM DESCRIPTION
F Air Inlet Fitting
A male DISS or NIST air fitting with one way valve
that is used to connect an air supply hose.
G Alarm
An audible alarm that sounds due to an excessive
pressure drop or deletion of either gas supply.
H Manifold Outlet (Optional)
Manifold with 3 primary outlets.
I Rear Slide Mount with dove tail.
INSTALLATION
WARNING
Read this User Manual before installing or operating the
Air-Oxygen Blender.
Confirm the concentration of air/oxygen with an Oxygen
Analyzer/Monitor.
1. Secure the Air-Oxygen Blender to a wall or pole bracket in an
upright position.
2. It is recommended to install a condensation trap in the air
supply line.
3. Connect the air and oxygen supply lines to the appropriate
inlet fittings on the bottom of the Blender.
4. Attach a flowmeter, or other metering device to one of the
outlet ports.
Primary Outlets Flow capacity:
• High Flow Blender (PM 5200 Model) 15 lpm to 120 lpm
• Low Flow Blender (PM 5300 Model) 3 lpm to 30 lpm
Auxiliary Outlet:
This will bleed off some of the air/oxygen mixture to
maintain concentration accuracy at the Low Flow setting.
High Flow Blender (PM 5200 Model) 15 lpm or less
• Low Flow Blender (PM 5300 Model) 3 lpm or less
5. Attach a supply line to the outlet port of the flowmeter.